What it does
This skill helps medical device teams run risk management under ISO 14971 across the life of a product. It is for risk managers and engineers who need documented, reviewed risk files.
When to use it
- ✓When you need to write a risk management plan for a medical device.
- ✓When you are doing hazard analysis, FMEA or fault tree work.
- ✓When you need to judge whether residual risk is acceptable or write a benefit-risk analysis.
Skill link
https://github.com/alirezarezvani/claude-skills/blob/main/ra-qm-team/skills/risk-management-specialist/SKILL.mdCopy it and paste it into your AI (Claude, ChatGPT or Cursor), asking it to follow the instructions.
How to use
The easiest way: copy the message and paste it into your AI. It reads the skill and follows the instructions.
Ready-to-paste message
For Claude Code, Codex or Cursor: run it in your project folder and the skill gets installed.
Download the SKILL.md and drop it into your agent’s skills folder.
Example of what to ask your AI
“Help me build the FMEA and risk matrix for our infusion pump's software and hardware hazards.”
5 new skills a week. 1 minute to read.
How it works
The agent sets risk acceptance criteria, guides the hazard analysis, evaluates risks and plans controls and their verification. It runs scripts for FMEA and risk matrix calculations.
Technical content (original SKILL.md)
Original text by the author, straight from GitHub.
Risk Management Specialist
ISO 14971:2019 risk management implementation throughout the medical device lifecycle.
Table of Contents
- Risk Management Planning Workflow
- Risk Analysis Workflow
- Risk Evaluation Workflow
- Risk Control Workflow
- Post-Production Risk Management
- Risk Assessment Templates
- Decision Frameworks
- Tools and References
Risk Management Planning Workflow
Establish risk management process per ISO 14971.
Workflow: Create Risk Management Plan
- Define scope of risk management activities:
- Medical device identification
- Lifecycle stages covered
- Applicable standards and regulations
- Establish risk acceptability criteria:
- Define probability categories (P1-P5)
- Define severity categories (S1-S5)
- Create risk matrix with acceptance thresholds
- Assign responsibilities:
- Risk management lead
- Subject matter experts
- Approval authorities
- Define verification activities:
- Methods for control verification
- Acceptance criteria
- Plan production and post-production activities:
- Information sources
- Review triggers
- Update procedures
- Obtain plan approval
- Establish risk management file
- Validation: Plan approved; acceptability criteria defined; responsibilities assigned; file established
Risk Management Plan Content
| Section | Content | Evidence |
|---|---|---|
| Scope | Device and lifecycle coverage | Scope statement |
| Criteria | Risk acceptability matrix | Risk matrix document |
| Responsibilities | Roles and authorities | RACI chart |
| Verification | Methods and acceptance | Verification plan |
| Production/Post-Production | Monitoring activities | Surveillance plan |
Risk Acceptability Matrix (5x5)
| Probability \ Severity | Negligible | Minor | Serious | Critical | Catastrophic |
|---|---|---|---|---|---|
| Frequent (P5) | Medium | High | High | Unacceptable | Unacceptable |
| Probable (P4) | Medium | Medium | High | High | Unacceptable |
| Occasional (P3) | Low | Medium | Medium | High | High |
| Remote (P2) | Low | Low | Medium | Medium | High |
| Improbable (P1) | Low | Low | Low | Medium | Medium |
Risk Level Actions
| Level | Acceptable | Action Required |
|---|---|---|
| Low | Yes | Document and accept; still reduce as far as possible (EU MDR) |
| Medium | After reduction AFAP | Reduce as far as possible; document why further reduction is impossible |
| High | After reduction AFAP | Reduction required; demonstrate all further options exhausted |
| Unacceptable | No | Design change mandatory |
EU MDR — AFAP, not ALARP: For CE-marked devices, risks must be reduced as far as possible (AFAP) without economic considerations (MDR Annex I, GSPR 1–4; EN ISO 14971:2019/A11:2021 Z-annexes deviation). ALARP ("as low as reasonably practicable"), which permits cost-benefit weighing in acceptability decisions, is not an acceptable criterion under the EU MDR — a notified body will flag it. ISO 14971:2019 itself removed ALARP from the normative text. ALARP may persist in some non-EU jurisdictions (e.g., the UK HSE tradition); if used outside the EU, flag the deviation from EU requirements explicitly.
Risk Analysis Workflow
Identify hazards and estimate risks systematically.
Workflow: Conduct Risk Analysis
- Define intended use and reasonably foreseeable misuse:
- Medical indication
- Patient population
- User population
- Use environment
- Select analysis method(s):
- FMEA for component/function analysis
- FTA for system-level analysis
- HAZOP for process deviations
- Use Error Analysis for user interaction
- Identify hazards by category:
- Energy hazards (electrical, mechanical, thermal)
- Biological hazards (bioburden, biocompatibility)
- Chemical hazards (residues, leachables)
- Operational hazards (software, use errors)
- Determine hazardous situations:
- Sequence of events
- Foreseeable misuse scenarios
- Single fault conditions
- Estimate probability of harm (P1-P5)
- Estimate severity of harm (S1-S5)
- Document in hazard analysis worksheet
FAQ
What standard does it follow?
It follows ISO 14971:2019 for medical device risk management.
Does it cover risks after a product is sold?
Yes. The skill includes post-production risk management.
Does it set up the risk matrix?
Yes. It sets probability and severity categories and a matrix with acceptance thresholds.
Details
Original description by the author
Use for ISO 14971 medical device risk management: hazard analysis, FMEA, fault trees, risk controls, benefit-risk or residual risk. Runs fmea_analyzer.py and risk_matrix_calculator.py.
Popularity
- GitHub stars
- 29k
- Ranking position
- #9742
- License
- MIT
- Updated
- 11/2/2026
Comments
- …
More skills from this repository
agent-launcher-orchestrator
29kHelps your agent plan, build, test and schedule a Claude managed agent in your own account, step by step, starting from one clear goal.
grade-iterate
29kLets your agent grade a managed AI agent's work against a rubric and improve it in a fixed number of rounds, so it never loops forever.
interview
29kLets your agent interview you about an AI agent idea and turn your answers into a simple build plan before anything is launched.
run-without-you
29kHelps your agent set up a managed AI agent to run on a schedule or when an event happens, with the cron timing checked before launch.
Similar skills
blucli
394kHelps your agent find, play, pause, group, and adjust the volume of Bluesound and NAD speakers through the blu command.
camsnap
394kHelps your agent grab snapshots, short clips, or motion events from network cameras and local webcams that you have set up.
eightctl
394kHelps your agent check and control an Eight Sleep pod, including its temperature, alarms, and sleep schedules.